1、INTERNATIONAL ATOMIC ENERGY AGENCYVIENNADigital mammography offers fundamental advantages over film based mammography.These include the possibility for acquiring quality images at lower radiation dose image recording,processing and archiving as well as the use of artificial intelligence for improvin
2、g diagnostic outcome.Other practical advantages include cost reduction,use of environmentally friendly technology,and the option of obtaining remote expert diagnostic opinion.Image quality in mammography is critical.A switch from screen-film technology to a digital system is preferable only if high
3、quality images can be guaranteed.This publication provides guidance on the establishment of digital mammography facilities and the upgrade of existing facilities.It focuses on planning,designing and operating the high quality mammography service within available resources.No.46Worldwide Implementati
4、on of DigitalMammography ImagingWorldwide Implementation of Digital Mammography ImagingIAEA HUMAN HEALTH SERIES No.46IAEA HUMAN HEALTH SERIESIAEA HUMAN HEALTH SERIESIAEA HUMAN HEALTH SERIES PUBLICATIONSThe mandate of the IAEA human health programme originates from Article II of its Statute,which sta
5、tes that the“Agency shall seek to accelerate and enlarge the contribution of atomic energy to peace,health and prosperity throughout the world”.The main objective of the human health programme is to enhance the capabilities of IAEA Member States in addressing issues related to the prevention,diagnos
6、is and treatment of health problems through the development and application of nuclear techniques,within a framework of quality assurance.Publications in the IAEA Human Health Series provide information in the areas of:radiation medicine,including diagnostic radiology,diagnostic and therapeutic nucl
7、ear medicine,and radiation therapy;dosimetry and medical radiation physics;and stable isotope techniques and other nuclear applications in nutrition.The publications have a broad readership and are aimed at medical practitioners,researchers and other professionals.International experts assist the IA
8、EA Secretariat in drafting and reviewing these publications.Some of the publications in this series may also be endorsed or co-sponsored by international organizations and professional societies active in the relevant fields.There are two categories of publications in this series:IAEA HUMAN HEALTH S
9、ERIESPublications in this category present analyses or provide information of an advisory nature,for example guidelines,codes and standards of practice,and quality assurance manuals.Monographs and high level educational material,such as graduate texts,are also published in this series.IAEA HUMAN HEA
10、LTH REPORTSHuman Health Reports complement information published in the IAEA Human Health Series in areas of radiation medicine,dosimetry and medical radiation physics,and nutrition.These publications include reports of technical meetings,the results of IAEA coordinated research projects,interim rep
11、orts on IAEA projects,and educational material compiled for IAEA training courses dealing with human health related subjects.In some cases,these reports may provide supporting material relating to publications issued in the IAEA Human Health Series.All of these publications can be downloaded cost fr
12、ee from the IAEA web site:http:/www.iaea.org/Publications/index.htmlFurther information is available from:Marketing and Sales UnitInternational Atomic Energy AgencyVienna International CentrePO Box 1001400 Vienna,AustriaReaders are invited to provide their impressions on these publications.Informati
13、on may be provided via the IAEA web site,by mail at the address given above,or by email to:Official.Mailiaea.org.WORLDWIDE IMPLEMENTATION OF DIGITAL MAMMOGRAPHY IMAGINGAFGHANISTANALBANIAALGERIAANGOLAANTIGUA AND BARBUDAARGENTINAARMENIAAUSTRALIAAUSTRIAAZERBAIJANBAHAMASBAHRAINBANGLADESHBARBADOSBELARUSB
14、ELGIUMBELIZEBENINBOLIVIA,PLURINATIONAL STATE OFBOSNIA AND HERZEGOVINABOTSWANABRAZILBRUNEI DARUSSALAMBULGARIABURKINA FASOBURUNDICABO VERDECAMBODIACAMEROONCANADACENTRAL AFRICAN REPUBLICCHADCHILECHINACOLOMBIACOMOROSCONGOCOSTA RICACTE DIVOIRECROATIACUBACYPRUSCZECH REPUBLICDEMOCRATIC REPUBLIC OF THE CONG
15、ODENMARKDJIBOUTIDOMINICADOMINICAN REPUBLICECUADOREGYPTEL SALVADORERITREAESTONIAESWATINIETHIOPIAFIJIFINLANDFRANCEGABONGAMBIAGEORGIAGERMANYGHANAGREECEGRENADAGUATEMALAGUYANAHAITIHOLY SEEHONDURASHUNGARYICELANDINDIAINDONESIAIRAN,ISLAMIC REPUBLIC OF IRAQIRELANDISRAELITALYJAMAICAJAPANJORDANKAZAKHSTANKENYAK
16、OREA,REPUBLIC OFKUWAITKYRGYZSTANLAO PEOPLES DEMOCRATICREPUBLICLATVIALEBANONLESOTHOLIBERIALIBYALIECHTENSTEINLITHUANIALUXEMBOURGMADAGASCARMALAWIMALAYSIAMALIMALTAMARSHALL ISLANDSMAURITANIAMAURITIUSMEXICOMONACOMONGOLIAMONTENEGROMOROCCOMOZAMBIQUEMYANMARNAMIBIANEPALNETHERLANDSNEW ZEALANDNICARAGUANIGERNIGE
17、RIANORTH MACEDONIANORWAYOMANPAKISTANPALAUPANAMAPAPUA NEW GUINEAPARAGUAYPERUPHILIPPINESPOLANDPORTUGALQATARREPUBLIC OF MOLDOVAROMANIARUSSIAN FEDERATIONRWANDASAINT KITTS AND NEVISSAINT LUCIASAINT VINCENT AND THE GRENADINESSAMOASAN MARINOSAUDI ARABIASENEGALSERBIASEYCHELLESSIERRA LEONESINGAPORESLOVAKIASL
18、OVENIASOUTH AFRICASPAINSRI LANKASUDANSWEDENSWITZERLANDSYRIAN ARAB REPUBLICTAJIKISTANTHAILANDTOGOTONGATRINIDAD AND TOBAGOTUNISIATRKİYETURKMENISTANUGANDAUKRAINEUNITED ARAB EMIRATESUNITED KINGDOM OF GREAT BRITAIN AND NORTHERN IRELANDUNITED REPUBLIC OF TANZANIAUNITED STATES OF AMERICAURUGUAYUZBEKISTANVA
19、NUATUVENEZUELA,BOLIVARIAN REPUBLIC OF VIET NAMYEMENZAMBIAZIMBABWEThe following States are Members of the International Atomic Energy Agency:The Agencys Statute was approved on 23 October 1956 by the Conference on the Statute of the IAEA held at United Nations Headquarters,New York;it entered into fo
20、rce on 29 July 1957.The Headquarters of the Agency are situated in Vienna.Its principal objective is“to accelerate and enlarge the contribution of atomic energy to peace,health and prosperity throughout the world.IAEA HUMAN HEALTH SERIES No.46WORLDWIDE IMPLEMENTATION OF DIGITAL MAMMOGRAPHY IMAGINGIN
21、TERNATIONAL ATOMIC ENERGY AGENCY VIENNA,2023 IAEA,2023Printed by the IAEA in AustriaOctober 2023STI/PUB/2011COPYRIGHT NOTICEAll IAEA scientific and technical publications are protected by the terms of the Universal Copyright Convention as adopted in 1952(Berne)and as revised in 1972(Paris).The copyr
22、ight has since been extended by the World Intellectual Property Organization(Geneva)to include electronic and virtual intellectual property.Permission to use whole or parts of texts contained in IAEA publications in printed or electronic form must be obtained and is usually subject to royalty agreem
23、ents.Proposals for non-commercial reproductions and translations are welcomed and considered on a case-by-case basis.Enquiries should be addressed to the IAEA Publishing Section at:Marketing and Sales Unit,Publishing SectionInternational Atomic Energy AgencyVienna International CentrePO Box 1001400
24、Vienna,Austriafax:+43 1 26007 22529tel.:+43 1 2600 22417email:sales.publicationsiaea.org www.iaea.org/publicationsIAEA Library Cataloguing in Publication DataNames:International Atomic Energy Agency.Title:Worldwide Implementation of Digital Mammography Imaging/International Atomic Energy Agency.Desc
25、ription:Vienna:International Atomic Energy Agency,2023.|Series:IAEA Human Health Series No.46,ISSN 20753772;no.46|Includes bibliographical references.Identifiers:IAEAL 23-01599|ISBN 9789201267238(paperback:alk.paper)|ISBN 9789201268235(pdf)|ISBN 9789201269232(epub)Subjects:Diagnostic imaging.|Diagno
26、stic imaging Methods.|Radiology.|Breast Radiography.|Radiography,Medical.Classification:UDC 615.849|STI/PUB/2011FOREWORDIn medical imaging,digital technologies can have many fundamental advantages over screen-film technology,such as(i)improved efficiency of the use of radiation in forming the image,
27、which allows for dose reduction,(ii)transmission of images electronically and from archival storage in a computer system,(iii)increased robustness to exposure techniques with the disappearance of obvious under-and overexposure failures and(iv)the many image processing and other digital applications,
28、such as artificial intelligence,that can be applied to digital images.Other practical advantages include cost reduction and replacement of chemical developers by a more environmentally friendly detector.Finally,the digital framework allows expert diagnostic judgements to be made available regardless
29、 of the distance between the imaging facility and the expert.These advantages also apply to mammography.However,in mammography,image quality is critical,and a sufficient level of quality is achieved only if system operation is optimized.A switch from screen-film to digital technology ought to be per
30、formed,therefore,only if high quality images can be guaranteed.Optimal system operation and quality control are needed at all levels including for the newest digital systems.It has been recognized that achieving optimized imaging in mammography is a complex multifactorial process that starts with th
31、e design and implementation of a proper infrastructure and the best technology.It requires well trained staff and a rigorous quality assurance programme.While many digital imaging devices exist,there is no unique answer to the best solution in practice,given set resources.It is also recognized that
32、there is no practical guidance on how to transition to digital technology systems.The World Health Organization(WHO)also recognized that a noteworthy number of clinical radiologists had experienced great challenges in implementing digital radiology and that radiology departments could benefit from a
33、n unbiased and independent resource to guide them in this.In 2016,the IAEA and WHO published IAEA Human Health Series No.28,entitled Worldwide Implementation of Digital Imaging in Radiology,a publication intended to address issues associated with the introduction of digital radiology.In response to
34、the need to advise Member States on topics specifically related to mammography,in 2016 the Scientific Committee of the IAEA/WHO Network of Secondary Standards Dosimetry Laboratories recommended that guidance be developed for the implementation of digital mammography technologies.The present publicat
35、ion is a companion to Human Health Series No.28 and provides information on the resources needed for different mammography systems.The IAEA acknowledges the contributions of the drafting committee responsible for the development of this publication in particular,H.Bosmans(Belgium),M.E.Brandan(Mexico
36、),R.A.Jong(Canada),M.Yaffe(Canada)and C.H.Yeong(Malaysia).The IAEA officers responsible for this publication were H.Delis and V.Tsapaki of the Division of Human Health.EDITORIAL NOTEAlthough great care has been taken to maintain the accuracy of information contained in this publication,neither the I
37、AEA nor its Member States assume any responsibility for consequences which may arise from its use.This publication does not address questions of responsibility,legal or otherwise,for acts or omissions on the part of any person.Guidance and recommendations provided here in relation to identified good
38、 practices represent expert opinion but are not made on the basis of a consensus of all Member States.The use of particular designations of countries or territories does not imply any judgement by the publisher,the IAEA,as to the legal status of such countries or territories,of their authorities and
39、 institutions or of the delimitation of their boundaries.The mention of names of specific companies or products(whether or not indicated as registered)does not imply any intention to infringe proprietary rights,nor should it be construed as an endorsement or recommendation on the part of the IAEA.Th
40、e IAEA has no responsibility for the persistence or accuracy of URLs for external or third party Internet web sites referred to in this book and does not guarantee that any content on such web sites is,or will remain,accurate or appropriate.CONTENTS1.INTRODUCTION.11.1.Background.11.2.Objective.11.3.
41、Scope.11.4.Structure.22.A ROAD MAP FOR A MAMMOGRAPHY FACILITY.32.1.Implementation scenarios.33.RESOURCES AND NEEDS.103.1.Project management.103.2.Core considerations.103.3.Specific needs for DDM.183.4.Specific needs for CRM.283.5.Specific needs for an SFM.413.6.Costing and implementation.504.TRANSIT
42、ION FROM SFM TO CRM AND DDM.584.1.General considerations.584.2.Radiation dose considerations.604.3.Impact of transition.624.4.Fixed installations versus mobile services.645.CONCLUSIONS.66APPENDIX I:BREAST CANCER 69APPENDIX II:MAMMOGRAPHY TECHNOLOGY 75APPENDIX III:OTHER BREAST IMAGING MODALITIES 92AP
43、PENDIX IV:MAMMOGRAPHY EQUIPMENT PERFORMANCE 96APPENDIX V:QUALITY MANAGEMENT 100APPENDIX VI:EDUCATION OF PROFESSIONALS IN MAMMOGRAPHY 110APPENDIX VII:EVALUATION OF MAMMOGRAPHY SERVICE 122REFERENCES.127ABBREVIATIONS.137GLOSSARY.139CONTRIBUTORS TO DRAFTING AND REVIEW.1411.INTRODUCTION1.1.BACKGROUNDThe
44、incidence of breast cancer is continuing to rise,particularly in low and middle-income countries(LMICs)1.Mammography is widely used to facilitate the detection of breast cancer at a point earlier in its natural history than is possible by clinical examination 2.It is also the only imaging method for
45、 the early detection of breast cancer that has been demonstrated to contribute to reduced mortality through screening 3.Its use is increasing in LMICs as the incidence of breast cancer increases.Digital mammography systems were introduced in 2000 and offer many advantages over screen-film mammograph
46、y.1.2.OBJECTIVEThe purpose of this publication is to provide guidance on the establishment of digital mammography facilities or upgrade of existing facilities when selecting and implementing new technology for mammography imaging in different resource settings.Guidance and recommendations provided h
47、ere in relation to identified good practices represent expert opinion but are not made on the basis of a consensus of all Member States.1.3.SCOPEThis IAEA publication is intended to assist health care policy decision makers,planners,programme administrators and professionals working in health care i
48、n establishing or upgrading capabilities for diagnostic and screening mammography.The IAEA and other organizations have published guidance around aspects related to breast cancer incidence 4,breast imaging 2,breast cancer prevention,quality assurance(QA)and quality control(QC)of screen-film 5 and di
49、gital systems 2,and QA in screening and diagnosis 6,7.The focus here is explicitly to inform and facilitate decisions and planning on how to select or transition mammography services to the best quality possible with available resources.11.4.STRUCTUREThe present IAEA publication builds on existing g
50、uidance 2,5,8 and presents a road map for how to move from any current situation to one that better addresses the needs of a community.Section 1 presents the background,objective,scope and structure of this publication.Section 2 shows a road map for a mammography facility and helps interested partie
51、s to identify the level of mammography service at which they are currently operating and the level that is feasible with upgraded services.Decision points in the road map are linked to relevant supporting information in subsequent sections and appendices.It also provides an overview of various imple
52、mentation scenarios with more detailed information to facilitate making decisions.Section 3 discusses what is required to implement the different types of service,direct digital mammography(DDM),computed radiography mammography(CRM)or screen-film mammography(SFM).Aspects of equipment,infrastructure,
53、quality needs,maintenance and staff training are included.In addition,the transition between screen-film and digital modalities is discussed in Section 4.More detailed information is provided in the appendices as follows:Appendix I provides a review on breast cancer incidence,mortality and geographi
54、cal factors as well as the principles of breast cancer screening and factors to consider before initiating a screening programme.Appendix II presents an overview of the different types of mammography technologies,including breast tomosynthesis and contrast-enhanced mammography.Appendix III reviews o
55、ther breast imaging modalities such as ultrasound,magnetic resonance imaging(MRI),positron emission tomography(PET)and breast computed tomography(CT).Appendix IV reviews evaluations of physico-technical parameters that describe the performance of mammography systems and presents examples of relation
56、ships between tests and clinical performance.Appendix V reports briefly on QA and QC tests that assure the high quality performance of systems,describes metrics for digital detectors and discusses the importance of artefact evaluation as well as the use of contrast detail phantoms to evaluate system
57、 performance.Appendix VI outlines the needs for professional education and training of key personnel involved in the delivery of high quality mammography services.Finally,Appendix VII shows some examples of existing mammography evaluation programmes implemented in some countries or continents.Ultima
58、tely,this publication potentially pertains to all people for whom the detection of breast cancer would be facilitated by mammography,but the term women is used because it reflects the vast majority of the use of breast imaging.22.A ROAD MAP FOR A MAMMOGRAPHY FACILITYThis section provides a road map
59、to assist health care decision makers,administrators and professionals.Establishment or upgrading of a mammography service is defined by applying various scenarios before deciding further steps.The analysis includes needs for investment and maintenance,as well as the availability of infrastructure a
60、nd human resources that are essential for evolution.Figures 16 are decision trees for various types of medical imaging facilities,and they have been created to guide the transition from a current situation towards improved mammography services.Cross-references to the relevant section in this publica
61、tion are provided to facilitate decisions.Although there are several technical and operational advantages associated with digital systems,it is also acknowledged that there may be differences in the available economic and human resources or infrastructure(e.g.quality of electrical supply)that may im
62、pact decisions that need to be made.Several scenarios exist and are explored to depict the next steps in improvement.If there is no scenario that exactly describes a facility,the two closest scenarios can be investigated.It is assumed as a minimum standard that,if mammography is performed,it is done
63、 using X ray equipment that is explicitly designed for the purposes of mammography in that it has the capability of producing low energy X rays(using different target/filter combinations),a small X ray focal spot,an automatic exposure control(AEC),an integrated compression device,and a mammography g
64、rid and beam collimator designed for imaging the breast.In addition,there ought to be a means available for providing the medical radiation technologist with information on the settings(kilovoltage(kV)and milliampere-seconds(mAs)for the exposure,where the AEC system cannot be used or is uncalibrated
65、.The system design and performance capability need to comply with the relevant current standards of the International Electrotechnical Commission(IEC)9.A general-purpose radiographic unit ought not to be used to carry out mammography.2.1.IMPLEMENTATION SCENARIOSDigital mammography plays a vital role
66、 in the diagnosis and management of breast diseases;however,its implementation needs to be carefully evaluated and justified according to the local scenario.This section discusses the practical implementation of digital mammography in multiple scenarios,such as(a)a facility that has limited resource
67、s and does not have any medical 3imaging equipment;(b)a facility that has limited resources and does not have a mammography system;(c)a hospital with an SFM facility;(d)a hospital with a CRM facility;and(e)a tertiary care hospital with breast imaging.2.1.1.No medical imaging facilityThis scenario is
68、 applied to a facility(e.g.a general clinic)that has limited resources(in terms of financial,human or infrastructure constraints)and does not have any medical imaging equipment yet.In this scenario,implementing other imaging capabilities such as general radiography or CT may be a higher priority tha
69、n mammography.A radiologist,medical radiation technologist and a clinically qualified medical physicist(CQMP)specifically trained in breast imaging are essential for introducing breast imaging.A facility lacking these professionals,even with general training,would have great difficulty establishing
70、high quality mammography.2.1.2.Limited resources no mammographyThis scenario refers to a setting where there is some general radiological imaging,but no mammography system.The term limited resources may refer to financial constraints,but in some cases,it denotes a lack of human resources 4FIG.1.Init
71、ial planning considerations for the implementation of mammography imaging.such as appropriately trained radiologists,medical radiation technologists,CQMPs or service engineers.If there is an identified need for breast imaging,then implementation of a breast imaging facility can be recommended.If scr
72、een-film general radiography is already being used in the facility,then a viable and realistic solution may be to consider initiating SFM.In addition to the need for purchasing an X ray mammography unit,however,dedicated cassettes,screens,films,mammography view boxes and a reading environment will b
73、e needed.A special processor will also be needed for developing mammography films.For chemical film processing,additional daily QC equipment and procedures will be required.Therefore,there are cost and time implications with the implementation of SFM.Furthermore,if no radiologist is available,it wou
74、ld also be necessary to transfer processed films to the location where the images would be reported.This could result in delays in obtaining a diagnosis.It may be more reasonable to initiate DDM in this scenario;or if adequate resources(for both purchase and maintenance)for DDM cannot be obtained,it
75、 is advisable to initiate CRM,although the latter is a less favourable choice than DDM.In all cases,specialized training for the medical radiation technologists on patient positioning,breast compression,exposure and QC will be necessary.In such a facility,the mammograms would likely mainly be used f
76、or diagnostic purposes rather than for screening asymptomatic women.If there is a radiologist available on-site,special training in interpreting mammograms will be required.Otherwise,remote interpretation via tele-mammography could be considered.In any scenario,QC,dose and image quality optimization
77、 as well as patient dose assessment and review ought to be implemented and overseen by a CQMP.Another solution,if appropriate,as part of a regional or national system,would be to consider participating in a mobile mammography programme 5FIG.2.Decision tree for facilities with established mammography
78、 imaging technologies.where a complete system,equipped with a DDM system and a trained medical radiation technologist,visits at regular intervals.Such a plan will also require a regular quality assurance programme(QAP)and optimization of practice for the equipment(see Section 4.4).2.1.3.Hospital wit
79、h an SFM facilityIn this scenario,the case of a local or regional hospital with existing general radiology and SFM systems is discussed.The way forward depends greatly on 6FIG.3.Decision tree for determining the appropriate mammography imaging technology.7FIG.4.Decision tree for a facility with SFM
80、considering an upgrade.FIG.5.Decision tree for a facility with CRM considering an upgrade.the quality of imaging the facility currently achieves,the resources available and the volume of work performed at the facility.If current mammographic imaging quality is compatible with recognized guidelines s
81、uch as those of the American Association of Physicists in Medicine(AAPM)10,American College of Radiology(ACR)11,European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services(EUREF)6 or the IAEA 2,5 and workload demands can be accommodated,then it is reasonable to conti
82、nue with SFM.If it is planned to convert existing general radiography equipment to digital,then a conversion of mammography to either a CRM or DDM system could be considered.If the mammography machine is due for replacement,purchase of a DDM system or,if not feasible,8FIG.6.Decision tree for facilit
83、ies considering a change to their DDM.a CRM system,is recommended.It ought to be mentioned that retrofits for digital detectors and acquisition systems are now available at a lower cost than for a DDM system and may be a viable option;these are preferable to computed radiography(CR)plates 12.It is e
84、ssential in this case that the basic mammography unit(e.g.X ray tube and generator,gantry,compression,etc.)be in appropriate operating condition.If the current quality of an SFM system is not adequate,this ought to ideally be addressed immediately.If image quality improvement is not possible or prac
85、tical,a switch to DDM may be the solution.2.1.4.Hospital with a CRM facilityFor a local or regional hospital with existing general radiology and CRM systems,the way forward will depend greatly on the current quality of imaging achieved as well as the economic resources available and the volume of wo
86、rk performed at the facility.Different scenarios and justifications are elaborated below:(a)The quality and efficiency of mammography could be improved by implementing DDM.If financial resources can be found,such a change is recommended.(b)If the current quality of CRM is not adequate,this problem o
87、ught to ideally be addressed immediately.If it is not possible or practical(e.g.without major purchases)to make these improvements with CR plate technology,it is recommended that the facility consider switching to DDM as part of the solution to achieving high quality.(c)If the current imaging qualit
88、y is compatible with recognized guidelines such as those of the AAPM 10,ACR 11,EUREF 6 or the IAEA 2 and workload demands can be accommodated,then it is reasonable to continue with the existing CRM system.If the mammography equipment is due for replacement,purchase of a DDM system is recommended.2.1
89、.5.Tertiary care hospital with breast imagingDDM is recommended in the case of a tertiary care hospital with a dedicated breast imaging service that includes multiple mammography units,several mammography medical radiation technologists,an adequate number of breast radiologists,and access to CQMPs a
90、nd qualified service engineers.Although DDM systems can offer the highest quality images from a technical point of view,this quality is possible only if the DDM system is operating properly and 9is being used correctly by appropriately trained personnel.Therefore,attention to QC and training standar
91、ds is essential.3.RESOURCES AND NEEDSIn this section,the resources and needs for different mammography technologies are described in detail.3.1.PROJECT MANAGEMENTFigure 7 shows the overall project management process in the establishment of a mammography facility.In developing the work plan,advice fr
92、om a professionally trained radiologist,medical radiation technologist and CQMP,with experience in mammography,would be invaluable.3.2.CORE CONSIDERATIONS3.2.1.Needs analysisNeeds analysis is the first step in project management.The needs analysis ought to be carried out on the basis of existing dem
93、ands as well as a realistic forecast.If an SFM is to be replaced by a DDM system,advice from experienced users would be useful.If local experience is not available,models from a similar health institution in the same country,or in a neighbouring country with similar socioeconomic structures,ought to
94、 be sought 8.The following points ought to be considered during needs analysis:Clinical need for breast imaging.For example,breast cancer is a significant problem,or its incidence is increasing rapidly in the region.Available clinical infrastructure for diagnosis and treatment of breast cancers.This
95、 includes the access to means of pathological diagnosis and access to treatments.Appropriate expertise to perform mammography diagnosis and screening.There ought to be an adequate number of qualified radiologists and medical radiation technologists to carry out routine mammography services.If local
96、10expertise(i.e.breast radiologist)is not available,telemammography may be considered.Financial resources to purchase the mammography equipment,support maintenance costs,replace expensive components at the end of their life and meet routine QC costs.11FIG.7.Overall project management process to esta
97、blish a mammography facility.Personnel to perform equipment maintenance,QC tests(daily tests by a medical radiation technologist and comprehensive tests by a CQMP)and professional advice and support from CQMP or information technology(IT)specialist.Expected patient volume or number of examinations p
98、er day.Possibility of lesion localization procedures.Possibility of advanced breast imaging techniques(e.g.tomosynthesis).Other breast imaging techniques(e.g.ultrasound)locally available.3.2.2.Facility designOnce the needs analysis is completed and needs are clearly identified,then discussion on the
99、 design of the facility will be initiated.More specifically,the following items can be investigated:Physical space:Based on the examination volume and the mix of procedures,the required amount of equipment and rooms can be determined.Personnel:Based on the workload,this includes breast radiologists,
100、medical radiation technologists,CQMPs and access to a qualified service engineer,administrative staff,IT support staff(if a digital system is chosen),etc.Workflow:The reception area,changing room,mammography equipment room,image processing area(e.g.darkroom for SFM,CR reader for CRM,digital workstat
101、ion,etc.),image display and radiologists reporting room,etc.ought to be located adjacent to each other for a better workflow.Supporting units such as IT and engineering services ought to be accessible or within the vicinity.Image and data management:Depending on the systems to be installed(SFM,CRM o
102、r DDM,either fixed or mobile),identify the storage requirements for images and data.Local regulations will dictate the minimum period of time they must be retained before they can be discarded.For digital data storage(cloud computing or a local server),it is necessary to comply with Health Level 7 s
103、tandards.1 Logistics:The mammography facility is usually located within an imaging department/centre so that the supporting services such as administration,scheduling,finance,etc.can be shared with other imaging services.1 http:/www.hl7.org123.2.3.Infrastructure and room When planning any medical X
104、ray imaging facility,it is necessary to ensure that nearby staff and public are not exposed to levels of radiation that exceed the current regulatory exposure limits 13.A mammography room may be smaller in size than other X ray rooms because of the smaller dimensions of the equipment.The room shield
105、ing requirements are also reduced due to the low X ray energies used.Because of this,normal building materials such as gypsum wallboard may provide sufficient attenuation.However,if this approach is used,it is important to remember that recalculation of the shielding will be required if the room is
106、used for other radiological purposes in the future 13.Depending on the design of the equipment,techniques used and total workload,radiation shielding requirements ought to be properly calculated by qualified personnel such as CQMPs or radiation protection experts and approved by the regulatory bodie
107、s 13.Details of the shielding calculation methods are described in Refs 1417.A typical mammography facility consists of the following structures(Fig.8):Reception area;Waiting areas(at least one for casual clothing and the other for gowned individuals);Changing room(s)either inside or adjacent to the
108、 mammography equipment room;Equipment room large enough for the mammography system and acquisition computer workstation(if applicable)with private access for the patient and staff;Darkroom(for SFM)or space for a CR plate reader(if applicable)close to the mammography unit;Radiologists interpretation
109、and reading room with appropriate viewing conditions(e.g.low ambient light level);Space allocated to additional service needs,if possible,such as medical physics,IT and engineering equipment;Spaces for professional training and meetings,if possible,with appropriate technology including computers,sof
110、tware,Internet access,online resources,etc.In addition,mammography equipment rooms require:Adequate power supply for uninterrupted function of the system without significant fluctuations.In all systems,an uninterruptable power supply(UPS)is recommended,while at least a simple surge protector ought t
111、o be 13added in the system when needed.Modern X ray generators can generally handle 5 second power outages,but not longer.Computers can be provided with a UPS,but normally the 10 to 30 minute range is the longest practical length of time for which these provide power.Temperature and humidity control
112、 to meet the needs of the imaging system.See the special needs of DDM systems in Section 3.3 both for image storage and operation.Dust control.Secure Internet access for software upgrades and possibly remote monitoring.Appropriate authorization(for the premises and the practices)according to the app
113、licable legislation,which includes radiation protection.3.2.4.EquipmentEquipment specifications need to be prepared by technical experts such as CQMPs,taking into consideration input from medical radiation technologists and radiologists,with full knowledge of the clinical needs and operational condi
114、tions,as well as regulatory requirements.Equipment specifications provided to the vendors ought to indicate the layout,the type of equipment/system needed,and the types of clinical procedures intended to be performed as well as a list of system components and description of the design,construction a
115、nd performance features of each component.Any electrical,mechanical and environmental conditions which may affect the performance of the equipment ought to also be included 13,18.14FIG.8.Sample mammography facility layout.The main components of a mammography system(Fig.9)include:X ray source(generat
116、or and tube)and gantry including compression paddle;X ray detector;Image processor(for an SFM or CRM system);Image display for interpretation;Data archive;Quality control tools and materials;UPS.As the designs and characteristics of the equipment depend strongly on the type of technology used,these
117、will be described in more detail in Sections 3.33.5.3.2.5.Authorization needsOnce the initial needs analysis is completed and discussion of the facility design is initiated,the authorization needs ought to be identified.These need to follow local or national legislation and regulatory bodies require
118、ments.This involves at least authorization by the regulatory body for radiation protection and by the health authority to carry out clinical practice.The authorization for radiation protection may be a two stage process of an initial application to build 15FIG.9.Main components of typical mammograph
119、y X ray equipment.a facility(submitted before construction begins)followed by a full review and assessment by the regulatory body,leading to granting of authorization 13.If the mammography system will be used for screening,additional justification and authorization might be needed.3.2.6.Staff Hiring
120、 and training of the dedicated staff to provide mammography services(including radiologists,medical radiation technologists and CQMPs)ought to be identified.In certain cases,when on-site staff cannot be available,access(e.g.through a contract)to adequate professional services is required.It is impor
121、tant to budget staff costs including training during the initial planning.3.2.7.Staff trainingIt is considered necessary for all staff members to have the certification required by local or national legislation to practice in the field of mammography.More information on the elements of the training
122、required for the imaging staff is discussed in Appendix VI.People with adequate training who are licensed to operate in a mammography facility ought to include the following staff:Radiologists;Medical radiation technologists;CQMPs;Administrative staff.Continuing education is important for radiology
123、professionals including those working in mammography.There are various methods(vendor videos,professional organizations,for-profit courses)to continually update knowledge in breast imaging.Most professional organizations recommend 15 hours of continuing education every three years,including the topi
124、c of radiation protection and medical exposure(justification and optimization).3.2.8.ProcurementOnce all of the above issues are addressed,then equipment procurement can be planned.Tender processes are usually applied in most centres to invite submission of quotations and technical specifications fr
125、om multiple vendors.The specifications need to be thoroughly reviewed to ensure that the qualified vendors have addressed all the identified needs of the equipment and facility.The quotation ought to also include a fully comprehensive maintenance and service 16contract,such as installation and calib
126、ration of the equipment,warrantees,delivery time,maintenance plans,QC tools,staff training and all other criteria as listed in the purchasers specifications.A purchase contract that sets out all items and conditions of the purchase as agreed by both the purchaser and the assigned vendor is recommend
127、ed.The contract ought to also include actions to be taken if conditions for acceptance are not met.A detailed and concise purchase contract will ensure the delivery of equipment in a timely and cost effective manner 18.3.2.9.Site-specific trainingThe training of all related staff on-site is vital fo
128、r the successful implementation of a mammography programme.This is particularly important for medical radiation technologists and radiologists.Plans for continuing education ought to also be identified.In developing the work plan,advice from professionally trained radiologists,medical radiation tech
129、nologists and CQMPs with experience in mammography would be valuable.The training programmes ought to include:Acceptance testing and commissioning of the equipment by a qualified CQMP;Implementation of a QAP;Breast screening and/or diagnostic workflow.3.2.10.Quality Every facility ought to develop a
130、nd put in place a relevant QAP with the participation of all professionals involved under the supervision of a CQMP.This is explained in more detail in Appendix V,but in terms of major quality needs,the QAP ought to include the following:(a)Preparation of relevant approval documents of the equipment
131、 for mammography use,as well as ensuring required approvals for electrical safety and radiation safety;(b)An organized QAP compatible with international or national recommendations that includes:(i)Acceptance test performed by a CQMP,with system adjustments until the system passes all the required p
132、hysico-technical tests.(ii)Approval of the clinical image quality by the radiologist team.(iii)Regular QC testing.Although it is very difficult to develop a technical QC test that correlates completely with clinical imaging performance,it is known that,to detect cancer,it is necessary to visualize s
133、ubtle 17changes in tissue for very small objects(e.g.microcalcifications)with a sufficiently high signal to noise ratio.It is desirable to do this at the lowest radiation dose compatible with such a task.There are various physical QC tests such as those developed and described as part of the program
134、mes from the AAPM 10,EUREF 6,the European Federation of Organisations for Medical Physics 19 and the IAEA 2,5.Tests of this sort ought to be used as part of the QC.Details of the testing are discussed in Appendix V.3.2.11.MaintenanceEvery facility ought to assure that maintenance is provided at leas
135、t yearly or in accordance with the recommendations of the manufacturer if that is more often 13.Maintenance for the first year after purchase is normally included as part of the purchase price and is called a warranty.Maintenance packages for subsequent services can be purchased after the end of the
136、 warranty.It is important for the facility to determine at the time of purchase precisely what is and what is not covered by such packages,as the details can vary tremendously.It is possible that details of these packages are negotiable,but this is usually done before the facility commits to purchas
137、ing the equipment.There are very expensive components,such as the X ray tube,X ray detector and high resolution monitors.If the cost of replacing such components is not included in the package,then funds to cover replacement cost need to be secured.Having a maintenance package may also guarantee a f
138、aster response by the service group in repairing system problems.Maintenance ought to be done by qualified personnel trained by the manufacturer.The equipment logbook ought to also be kept up to date with all maintenance details and notes.3.3.SPECIFIC NEEDS FOR DDMThe specific needs for a DDM facili
139、ty are discussed in this section.Figure 10 shows the typical workflow in a DDM facility.With the advantages of digital technology,the mammographic image and data can be viewed and analysed immediately once acquired,conveniently stored,printed and transmitted to the picture archiving and communicatio
140、n system(PACS).183.3.1.DDM equipmentImportantly,all mammography equipment ought to comply with relevant standards for mammography 9 as minimum requirements to ensure suitability for clinical use.A clinical DDM system includes the following:X ray source(X ray generator and X ray tube);Digital detecto
141、r;Acquisition workstation/operating console;UPS;Image processing and diagnosis workstation;Applicable software.Table 1 shows the recommended minimum specifications for a modern DDM system.Optionally,the system could be upgradeable to one with contrast-enhanced digital mammography(CEDM)and digital br
142、east tomosynthesis(DBT).Figure 11 shows the typical layout of a DDM equipment room while Fig.12 illustrates the main components of DDM X ray equipment.19FIG.10.Typical workflow in a DDM facility.Once the digital image is acquired,it can be shared instantly with the image processing software(puter ai
143、ded diagnosis),radiologists workstation and PACS server for evaluation.The image can either be stored digitally or printed in hard copy.(Images courtesy of Hospital Sungai Buloh,University of Malaya Medical Centre and Bin Zheng,Malaysia.)20TABLE 1.RECOMMENDED MINIMUM SPECIFICATIONS FOR A MODERN DDM
144、SYSTEM 2,6,11,2022 X ray generatorPower5.0 kWTube voltage range 2435 kVp(maximum 1 kVp step)Tube load rangeAt least 5400 mAsAECManual and automated selection of kVp,mAs,filterExposure time rangeAt least 30 ms to 2 sAnode and filter materialMaterials need to allow for low dose/high penetration spectr
145、a,even for thick or dense breasts.Examples of appropriate target materials are,for X rays:molybdenum(Mo),rhodium(Rh),tungsten(W)and for beam filter(s):Mo,Rh,aluminium(Al),silver(Ag)Focal spotTwo(approximately 0.3 mm and 0.1 mm);with both manual and automated selection of the focal spot availableAnod
146、e heat capacityAt least 200 000 HUX ray tube and gantryCollimators Fully automated adjustment for different paddles,sizes and magnificationMovementsMotorized vertical and rotating movementArm locking systemElectromechanical brakes or equivalentArm moving systemMotorizedControl buttons for vertical a
147、nd rotational movementOn both sides of C-armFocal spot to image detection distance(SID)65 cmPatient face shieldProvided for 2-D imaging21TABLE 1.RECOMMENDED MINIMUM SPECIFICATIONS FOR A MODERN DDM SYSTEM 2,6,11,2022 (cont.)Breast compression Manual and automated breast compression Maximum force when
148、 automated is between 150 N and 200 N Emergency release Automatic decompression after exposureDigital numerical indicator,both sides For C-arm rotation angle and compression forceMemorizable mediolateral oblique(MLO)angleC-arm that can stop automatically at contralateral angleScatter rejectionAntisc
149、atter grid(or equivalent technology)Magnification views At least one magnification view(with magnification stand)Magnification ratio provided,display providedAntiscatter grid removalAutomatic,motorized for magnification viewsBreast compression paddlesSeveral sizes(e.g.18 cm 24 cm and 24 cm 30 cm)wit
150、h small breast,spot and magnification paddles includedDigital detectorDetector typePreferably direct conversion flat panel detector(amorphous selenium(a-Se)or similar)Effective field sizeAt least equivalent to 24 cm 29 cmPixel size100 mDead pixel mapProvidedImage depthAt least 12 bitSystem spatial r
151、esolution7 lp/mmAcquisition workstationGeneral A separate workstation for image positioning and patient demographic data22TABLE 1.RECOMMENDED MINIMUM SPECIFICATIONS FOR A MODERN DDM SYSTEM 2,6,11,2022 (cont.)Operator controlsBoth detector and generator controls integrated in the same consoleExposure
152、 parameters settingBoth manual and fully automaticStandard clinical protocolsDocumented on a chart and/or preprogrammed within the systemComputer systemLatest technology(processor generation/type/speed,RAM or operating memory,hard disk,storage systems,etc.)Storage capacityAccording to the facilitys
153、need(e.g.minimum 5000 patients,both projections,both breasts)MonitorAt least 3 MP high resolution flat panelTime between sequential acquisitions30 sTime between acquisition and workstation preview60 cm31TABLE 3.RECOMMENDED MINIMUM SPECIFICATIONS FOR A CRM SYSTEM 2,28,29 (cont.)Patient face shieldPro
154、videdBreast compression Manual and automated breast compression Maximum automated force between 150 N and 200 N Emergency release Automatic decompression after exposureDigital numerical both sides indicatorFor C-arm rotation angle and compression forceMemorizable MLO angleC-arm that can stop automat
155、ically at a contralateral angleScatter rejectionAntiscatter grid(or equivalent technology)Magnification viewsAt least one magnification view(with magnification stand)as specified;magnification ratio needs to be providedAntiscatter grid removalAutomatic,motorized for magnification viewsBreast compres
156、sion paddlesSeveral sizes available(e.g.18 cm 24 cm and 24 cm 30 cm)with small breast,spot and magnification paddles includedCassette holder and buckyAvailable to hold different cassette sizes(i.e.18 cm 24 cm and 24 cm 30 cm)and a large bucky if appropriate to the populationAcquisition workstationGe
157、neral A separate workstation for image positioning and patient demographic dataOperator controlsGenerator controls integrated in the same consoleExposure parameters settingBoth manual and fully automaticStandard clinical protocolsDocumented on a chart and/or preprogrammed within the system32TABLE 3.
158、RECOMMENDED MINIMUM SPECIFICATIONS FOR A CRM SYSTEM 2,28,29 (cont.)Computer systemLatest technology(processor generation/type/speed,RAM,hard disk,storage systems,etc.)Storage capacityAccording to the facilitys needs(e.g.minimum 5000 patients,both projections,both breasts)MonitorAt least 3 MP high re
159、solution flat panelDICOM functionality and connectivityDICOM compatible(e.g.DICOM 3.0)Image processingImage analysis tools(e.g.mean,standard deviation,and region of interest tool with adjustable size and shape)Additional capabilities Ability to perform repeat/reject analyses Special processing for i
160、mplants Ability to add comments to an imageHand/foot switchesDouble hand switch for the exposure and double foot switch for breast compression/arm movements Radiation protection barrier dimensionsAdequate for operator protection(as calculated by the qualified personnel)Dimensions need to be provided
161、Radiation protection barrier thickness Adequate thickness to protect the operator(as calculated by the qualified personnel)Lead equivalence(mmPb)needs to be providedComputed radiography detector(imaging plate)Detector typeReusable photostimulable storage phosphor imaging plate stored in a cassetteEf
162、fective field size Commonly available in two sizes:18 cm 24 cm and 24 cm 30 cmPixel size50 mImage depthAt least 12 bit33TABLE 3.RECOMMENDED MINIMUM SPECIFICATIONS FOR A CRM SYSTEM 2,28,29 (cont.)System spatial resolution10 lp/mmNumber of imaging plate recommendedAt least two of each size;preferably
163、at least four of each sizeComputed radiography readerLaser spot size 50 m(modern reader)100 m(conventional reader)Readout time 50 s for a 18 cm 24 cm imaging plate 60 s for a 24 cm 30 cm imaging plateTransition time for digital image to be displayed on the workstation3060 sErasure time1020 sTotal cy
164、cle time(insert,read,remove and insert next cassette)6090 s(depending on the imaging plate size and signal density on the plate)Image processing and diagnosis workstationWorkstation System of the latest technology workstation class hardware(processor generation/type/speed,RAM,hard disk,storage syste
165、ms,etc.)MonitorsAt least two approved diagnostic mammography monitors of at least 5 MP and 53.5 cm(21 inches),with automated self-calibrationDisplay graphic cardHigh end medical gradeStorage capacityAs required(e.g.at least 1.5 TB)Workstation capabilities Display of multiple images and priors for co
166、mparison purposesMultimodality viewer capability for display of ultrasound,X ray,digital mammography,MRI,PET,CT on a third colour monitor34TABLE 3.RECOMMENDED MINIMUM SPECIFICATIONS FOR A CRM SYSTEM 2,28,29 (cont.)Mouse,keypadDedicated keypad for mammography,common keypad,mouseSoftware and interconn
167、ectivitySoftwareDedicated breast imaging software with at least the following functions:Magnification Zoom Pan Windowing Brightness adjustment Contrast adjustment Distance measurement Histogram display Contrast enhancementDisplay capabilitiesOne-to-one image display:One pixel on detector correspondi
168、ng to one pixel on monitorInterconnectivity Full DICOM compatibility(e.g.DICOM 3.0);Media export/import Automated transfer of the image acquisition parameters(e.g.kV,target and filter material,breast thickness,compression force,projection,L/R)for each exposure into the DICOM header Capability to exp
169、ort unprocessed and processed DICOM imagesPower supplyUPSDedicated on-line UPS(for all workstations and accessories)for a minimum backup time of approximately 30 minutesFigure 16 shows the main components in a CRM equipment room and Fig.17 illustrates the structure of a CRM cassette/imaging plate.35
170、36FIG.16.Typical layout of a CRM equipment room.(Image courtesy of CPD Projects,Australia.)FIG.17.Inner structure of a mammography CR cassette.3.4.2.Computed radiography readerThe CR reader can be used exclusively for mammography imaging plates or can be shared with other radiography plates.Ideally
171、it needs to be optimized for mammography use with the specific imaging plate technology that is present.The readers are available in different designs(Fig.18)to suit the local needs.The reader needs to be equipped with a computer and a display to register the patient,identify the plate and review or
172、 check the image.It ought to be capable of dedicated image processing for a breast and of producing DICOM compatible image formats.It is desirable for technical data such as exposure parameters,compression force and mean glandular dose(MGD)to be automatically transferred to the appropriate tags in t
173、he DICOM header to allow tracking.Alternatively,these can be manually registered,although this can be time consuming.The reader needs to be connected to PACS or to another storage device,whereby it ought to permit the export of images to other digital media,and appropriate software licences to do th
174、is need to be in place.It ought to be possible to export DICOM for processing images for QC purposes too.3.4.3.Review workstationThe review workstation requirements for a CRM facility are in general similar to those of a DDM 2.Also in CRM,two high resolution monitors of at least 5 MP are required.If
175、 it is not possible to purchase two monitors,software ought to be available to allow side by side viewing of images on a single high 37FIG.18.Samples CR readers:(a)tabletop;(b)single slot;and(c)multiple slots.resolution monitor.As with DDM,the monitor(s)are calibrated according to the Grayscale Stan
176、dard Display Function and tested regularly 2,25 using the standard Society of Motion Picture and Television Engineers test pattern(Fig.19)25.3.4.4.Data archivingIdeally,the system will be connected to a PACS system in the facility.Otherwise,some other form of adequate digital storage is required tha
177、t is appropriate for the workload of the facility.Approximately 40 MB is required for each examination that uses CR plates,assuming lossless compression.38FIG.19.Standardized Society of Motion Picture and Television Engineers test pattern,one type of test pattern that may be provided with the medica
178、l diagnostic monitor for quality control.3.4.5.Image printingIdeally,image interpretation will be done on a workstation with high resolution monitors so that the image can be manipulated interactively while viewing.If this is not feasible,images can be printed on a high resolution film printer.This
179、ought to be done at full size(1:1,one pixel printed for one pixel in the image)so that information is not lost.If interpretation is done from films(printed images),suitable film viewers(view boxes)are required.3.4.6.Quality The quality needs are as stated in Section 3.2.10,with the following additio
180、ns:A comprehensive QAP for CRM needs to be in place,as described in international guidelines 2,6,7.Test objects,software tools and dosimetry equipment ought to be available for QC of the CRM system.It is recommended to include such equipment in the procurement plan of new installations to minimize o
181、r optimize costs and avoid subsequent unanticipated expenses.System optimization to detect small,subtle structures in the breast.Because CRM is less dose efficient than DDM,the main task is to get the best performance it can deliver.This is often quantified by achieving at least a specified value of
182、 the signal difference to noise ratio(SDNR)as a ratio to dose.3.4.7.MaintenanceAlthough CR plates are part of the digital imaging component of the system,they are considered consumables and they need to be replaced when their performance is suspected to have deteriorated below set quality limits.Thi
183、s is a cost that needs to be foreseen in the planning of new practices and will depend on the workload of the department.The lifespan of a CR plate is typically hundreds or thousands of exposurereadouterase cycles,determined primarily by the care in handling and environmental conditions.During their
184、 lifespan it is important that the CR cassettes and screens be kept in good working condition,and one staff member(usually the medical radiation technologist)ought to have the responsibility for their regular maintenance,such as cleaning,removing from service and requesting replacement of those cass
185、ettes that have degraded or cannot be cleaned.Maintenance ought to be done by personnel qualified and trained by the manufacturer.Maintenance logbooks need to be kept up to date.393.4.8.Staff trainingTransition from film to CRM is a challenging task and requires adequate competencies for all involve
186、d professionals.Some of the required additional skills for radiologists,medical radiation technologists and CQMPs are shown in Table 4.TABLE 4.ADDITIONAL SKILLS REQUIRED IN CRM FOR A RADIOLOGIST,MEDICAL RADIATION TECHNOLOGIST AND CQMP PersonnelAdditional skillsRadiologist Basic principles of CRM Bas
187、ic understanding of the modality and technique Reading/reviewing the modality(or telemammography)Basic computer skills especially related to digital image postprocessing Accommodation to the different,possibly unfamiliar appearance of digital images Judgement of clinical image quality In-depth train
188、ing on optimal use of the modality,including image presentation,storage and retriaround the breast to acquire multiple projections.These projections are then reconstructed to produce a series of images in closely spaced focal planes parall